Delloyd's Lab-Tech Chemistry
LABGUIDE
	 
		
	Determination of Zinc by Atomic Absorption spectrometry
 
 
  
Equipment: 	Atomic absorption Spectrophotometer 
 
Sample:	Three samples of known concentration given. (See Solutions Preparations below) 
 
Apparatus:
  	10ml or 5ml burettes; grad pipettes 1ml, 2ml, 5ml; volumetric flasks 100ml; burette clamps and stands; pH 4 test paper; indicator bottles. 
 
Chemicals:
  	EDTA (ethylene diamine tetra acetic acid) , or (ethylenedinitro) tetra acetic acid). 
(HO2CCH2)2NCH2CH2N(CH2CO2H)2,  FW = 292.24, 
Zn metal, 
Conc. HCl 
Xylenol orange (indicator), Dye content = 90% 
Hexamine (solid) - Hexamethylenetetramine  FW = 140.19 
 
Solutions preparation:		 
 
1. 0.1M  EDTA standard soln. Dry salt in oven at 80°C, weigh out 37.225g and make up to 1 liter with distilled water.  
2. 1000ppm Zn standard. Dissolve 1g pure Zn metal in 30ml of 5M hydrochloric acid, and make up to 1 liter with distilled water. 
3. Provide three Zn solutions of 40ppm, 60ppm,and 80ppm, labeled A, B, and C, respectively. 	 
4. 40ppm = 40mls of 1000ppm made up to 1 liter = A 
      60ppm = 60mls of 1000ppm made up to 1 liter = B 
      80ppm = 80mls of 1000ppm made up to 1 liter = C 
5. 0.5% xylenol orange indicator (dissolve 0.5g in 100ml distilled water) 
6. Hexamine. (provide solid) 
 
Analytical notes: 
 
1. Provide 100ppm Zn standard for AA measurements (Dilute 100ml of the 1000ppm to 1 liter solution, or 10ml to 100 ml vol.) 
2. Range for AA is 0.1ppm to 2.0ppm stds.  (0.1, 0.5, 1.0, 1.5, and 2.0ppm) 
3. The dilution factor for EDTA is 1 in 20 (0.005M) 
4. The Zn solns. provided must be acidic. To ensure this, add 10mls conc. HCl to the preparation of each solution. (pH < 1) 
5. If there is no color change for blank, there is no zinc: Use blank value as zero.
  
Operation of Instrument: 
 
Follow Users Guide. Obtain help and advice from Demonstrator or Technician
 
  
 