A labguide has been produced for each experiment. It gives info on
preparations, chemicals, apparatus and sample requirements. It also
includes a helpful analytical notes feature.
To get the labguide for this experiment click here Enjoy!
Revised labscript :
Investigation of some factors affecting accuracy of sodium determination
and their application in the determination of sodium
by Flame Emission Spectroscopy
1. Introduction:
Atomic emission spectroscopy (AES) employing flames, also called flame emission spectroscopy (FES) or flame photometry has found widespread application in elemental analysis (1). Its most important uses have been in the determination of sodium, potassium, lithium and calcium, particularly in biological fluids and tissues. For reasons of convenience, speed, and relative freedom from interferences, flame emission spectroscopy has become the method of choice for these otherwise difficult to determine elements. The method has also been applied, with varying degree of success, to determine of perhaps half the elements in the periodic table.
Flame photometry is based on the emission spectrum of an element which is excited in a flame (e.g. propane/air, acetylene/air) which is hot enough to cause the element to allow emission of characteristic radiation. The spectrum may be relatively simple, consisting of only a few lines, or may be complex, broad bands.
Measurement of the intensity of a portion of a spectrum characteristic of an element can provides a measure of the concentration in the sample. The required spectral line or a portion of the spectrum is isolated either by a monochromator or by an optical filter. The intensity of the isolated radiation is measured by a photosensitive detector coupled to an amplifier and recorder.
In the analysis of sodium and potassium in the presence of calcium, some interference by the latter occurs, due to spectral overlap. The addition of a sufficient quantity of aluminium ions to the analyte solution tends to reduce the emission due to calcium and hence minimize the interference (2, 3).
Potassium determination by flame emission is affected mainly by ionization of potassium at the high temperatures associated with air/acetylene or hotter flames, especially at low concentrations of the elements. However this effect is negligible in the air/propane flame used by the flame analyser in this experiment. Therefore, addition of radiation buffers is not required for potassium analyses with this instrument.
2. Experimental procedures:
(A) Optimization of the fuel/air flow rates for determination of sodium:
(a) Prepare a calibration series containing 1.25, 2.5, 5.0 and 7.5ug/mL sodium
respectively in distilled water (25mL each), using the 50ug/mL standard sodium
solution provided.
(b) Set the sodium filter in position before the photocell and set the air pressure to the
burner as recommended (~20 psi)
(c) Depress ignition switch to light the flame and slowly increase the fuel flow rate. Once
the flame is lit, observe the flame, and carefully adjust the fuel flow rate, until a non-
luminous flame is obtained. Allow the system to equilibrate for about 15min.
(d) Aspirate the 2.5ug/mL sodium standard and adjust the sensitivity knob to obtain an
emission reading of about 30 units. Re-zero the instrument while aspirating distilled
water.
(e) Aspirate again the 2.5ug/mL solution, and carefully change the fuel flow rate until a
maximum signal is obtained. Avoid using a luminous flame, which creates a high
background signal.
(f) Re-zero with distilled water again. The instrument is now ready for use.
(B) Investigation of effect of aspiration rate on emission signals of sodium:
(a) Into each of four 25mL volumetric flasks, pipette the required volume of standard
100ug/mL sodium standard, to give a final concentration of 2.5ug/mL.
(b) Add to the flasks 2.5, 5.0, 7.5 and 10mL respectively of ethanol and make up to the
mark with distilled water. These correspond to 10 to 40% by volume of ethanol in
sodium solutions. Similarly prepare a series of 10-40% ethanol blank solutions in
distilled water.
(c) Using distilled water, zero the instrument and then aspirate the 2.5ug/mL sodium
standard in distilled water (0% EtOH), noting its emission value.
(d) Aspirate and measure the emission intensities of the distilled water sodium standards
(1.25 to 7.5ug/mL).
(e) Aspirate the ethanol samples in order of increasing ethanol content, and note the
corresponding readings.
(f) Plot emission vs ethanol (0-40% v/v) content of the solution, corrected for blanks.
(g) Exercise: What precautions should be taken when determining sodium in alcoholic
beverages?
(C) Interference of sodium emission by calcium and the countering effect
of aluminium:
(a) Into five 25mL volumetric flasks, add the required volume of the 100ug/mL sodium
standard stock solution that will give respective final concentration values of 1.25, 2.5,
5.0, and 7.5ug/mL.
(b) To each flask, add 10mL of the 1000ug/mL calcium standard stock solution provided
and make up to the mark with distilled water.
(c) Prepare a blank solution containing 10mL of the 1000 ppm calcium stock solution in
25mL solution.
(d) Into five other 25mL volumetric flasks, add similar quantities of sodium and calcium
solutions. Then add to each, 5mL of the aluminium solution provided and make up to
the mark with distilled water.
(e) Prepare a blank containing 10mL of 1000ug/mL calcium standard stock solution and
5mL of the aluminium solution in 25 mL solution.
(f) Aspirate in the following order:
(i) The standard solutions in distilled water (1.25-7.5ug/mL), with distilled water as
blank.
(ii) The sodium-calcium solutions (1.25-7.5ug/mL)and blank.
(iii) The sodium-calcium-aluminium solutions (1.25-7.5ug/mL) and blank.
(iv) Sample solutions and blank (preparation in later section).
(v) The sodium-ethanol solutions and blanks.
(vi) Tabulate the emission readings. Note the change of color of the flame in the three
types of solutions.
3. Exercises:
(A) Plot blank-corrected emission values vs sodium concentration of each set of solutions
(i), (ii) and (iii) on the same sheet of graph paper, or using a plotter.
Comment on the coincidence of the calibration curves, and the effectiveness of the
Al+3 in suppressing Ca+2 interference.
(B) Plot the emission values for the sodium-ethanol solutions, corrected for ethanol-
water blanks, vs EtOH concentration.
(a) Discuss the effect of viscosity and fuel/air ratio changes respectively, on the
instrumental response to a fixed sodium concentration.
(b) How would you determine the concentration of sodium in an alcoholic beverage?
4. Analysis of sodium in sample:
(a) Place triplicate, accurately weighed or measured quantities of the
sample provided into separate boiling tubes.
(b) To each tube add 5mL conc. nitric acid and reflux at 135-140°C
for 1h. in a fume hood.
(c) Prepare an acid blank simultaneously
(d) Cool and dilute with 10mL distilled water
(e) Filter into 100mL volumetric flasks and make to volume with distilled
water rinses of the boiling tubes and filter papers.
(f) Aspirate your sample solutions along with the solutions of section
and determine the mean sodium content of the samples.
5. References:
1. Flame Emission and Atomic Absorption Spectroscopy, Vols. 1, 11, 111. Dean, J.A.,
Rains, T.C. Eds. Marcel Decker, NY. 1969-75
2. Analyst 77 430-436 (1952)
3. Analyst 82 200 (1957)
anal-chem resources Chem. Dept. UWI. St. Augustine Campus
Original labscript :
Determination of sodium by Flame photometry
Introduction
1. Optimization of fuel-air ratio for sodium determination by Flame
emission spectroscopy
2. Investigation of some factors affecting accuracy of sodium
determination
3. Analysis of sodium in sample
Atomic emission spectroscopy (AES) employing flames, also called flame
emission spectroscopy (FES) or flame photometry has found widespread
application in elemental analysis. Its most important uses have been in
the determination of sodium, potassium, lithium and calcium,
particularly in biological fluids and tissues. For reasons of
convenience, speed, and relative freedom from interferences, flame
emission spectroscopy has become the method of choice for these
otherwise difficult to determine elements. The method has also been
applied, with varying degree of success, to determine of perhaps half
the elements in the periodic table.
Theory
Flame photometry is an analytical technique based on the spectrum of an
element when a solution containing it is aspirated into a flame (e.g.
propane/air, acetylene/air) which is hot enough to cause the element to
emit its characteristic radiation. The spectrum is normally relatively
simple, consisting of only a few lines, and measurement of the
intensity of one or more of these provides a highly sensitive measure
of the concentration in the sample. When several elements are present,
the required spectral line is isolated either by passing the light
from the flame into the entrance slit of a monochromator by means of a
lens or mirror or by a system of optical filters. The intensity of the
isolated radiation is measured by a photosensitive detector coupled to
an amplifier and recorder.
In the analysis of sodium and potassium in the presence of calcium,
some interference by the latter due to spectral overlap occurs. The
addition of a sufficient quantity of aluminium ions to the analyte
solution tends to reduce the emission due to calcium and hence minimize
the interference.
Potassium determination by flame emission is affected mainly by
ionization of potassium at the high temperatures associated with
air/acetylene or hotter flames, especially at low concentrations of the
elements. This effect is however, negligible for the air/propane flame as
used by the flame analyser in this experiment. Therefore, addition of
radiation buffers is not required.
Procedure
A. Optimization of the fuel/air flow rate for determination of
sodium
(1) Prepare a calibration series containing 1.25, 2.5, 5.0, and 10 ppm
sodium in distilled water (50 mls each) from the 50 ppm standard stock
sodium solution provided.
(3) Set the sodium filter in position before the photocell and set the
air pressure to the burner as recommended (~20 psi)
(4) Depress ignition switch to light the flame and slowly and slowly
increase the fuel flow rate. Once the flame is lit, observe the flame,
and carefully adjust the fuel flow rate until a non-luminous flame is
obtained. Allow the temperature to equilibrate for about 3-5
minutes.
(5) Aspirate the 2.5 ppm sodium standard and adjust the sensitivity
knob to obtain about an emission reading of about 0.3. Re-zero the
instrument while aspirating distilled water
(6) Aspirate again the 2.5 ppm solution, and carefully change the fuel
flow rate until a maximum signal is obtained, but avoid using a
luminous flame.
(7) Re-zero with distill water again. The instrument is now ready for
use.
B. The effect of aspiration rate on sensitivity of detection of
sodium
(1) Into each of four 25 ml volumetric flasks, pipette the required
volume of standard 100 ppm sodium standard to give a final
concentration 2.5 ppm. Add to the flasks 2.5, 5.0, 7.5, and 10.0 mls
respectively of methanol and make up to the mark with deionized
distilled water.
(2) Aspirate the 2.5 ppm sodium made up in distilled water only and
note the emission reading.
(3) Using deionised distilled water, zero the instrument and then
determine the emission of the distilled water standards.
(4) Aspirate the methanol samples in order of increasing methanol
content, and note the corresponding readings.
(5) Plot emission vs methanol (0-40%) content of the solution.
What precautions should be taken when determining sodium in alcoholic
beverages?
C. Interference of sodium emission by calcium and the countering
effect of aluminium
(1) Into five 25 ml volumetric flasks, add the required volume of the
100 ppm sodium standard stock solution that will give respective final
concentration values of 1.25, 2.5, 5.0, and 10 ppm. To each flask add
10 ml of the 1000 ppm calcium standard stock solution provided and make
up to the mark with deionized distilled water.
(2) Prepare a blank solution containing 10 ml of the 1000 ppm calcium
stock solution in 25 ml volume solution.
(3) Into five other 25 ml volumetric flasks, add similar quantities of
sodium and calcium solutions. Then add to each 5 ml of the aluminium
solution provided and make uo to the mark with distilled water.
(4) Prepare a blank containing 10 ml 1000 ppm calcium standard stock
solution and 5 ml of the aluminium solution in 25 ml volume
solution.
(5) Aspirate the standard solutions in distilled water, with distilled
water as blank, then the sodium-calcium solutions, and then the
sodiun-calcium-aluminium solutions, and note the emission/sodium
concentration readings.
(6) Note and comment on the change of color of the flame.
(7) Aspirate the respective blanks and subtract their emission values
from the sample solution values to provide blank-corrected emission
values.
(8) Plot corrected emission values vs sodium concentration of each set
of solutions on the same sheet of graph paper, or using a
plotter.
Comment on coincidence of the calibration curves, and
the effectiveness of the Al+3 in suppressing Ca+2
interference.
D. Analysis of sodium in sample
(1) Place triplicate, accurately weighed or measured quantities of the
sample provided into separate boiling tubes
(2) To each tube add 5 ml pure nitric acid and allow to predigest for
at least one hour in a fume hood.
(3) Prepare an acid blank simultaneously
(4) Reflux the material gently on dry heating block for one hour in a
fume hood, ensuring that fumes are properly vented.
(5) Cool and dilute with 10 ml distilled water
(6) Filter through a Whatman filter paper No. 1 into 100 ml volumetric
flasks and rinse the boiling tube and filter paper with 25 ml distilled
water.
(8) Aspirate your sample solutions along with the solutions of section
D, (5) and determine the mean sodium content of the samples.
References
1. Flame Emission and Atomic Absorption Spestroscopy, Vols 1, 11, 111.
Dean, J.A., Rains,T.C. Eds. Marcel Decker, NY. 1969-75
2. Analyst 77 430-436 (1952)
3. Analyst 82 200 (1957)
anal-chem resources Chem. Dept. UWI. St. Augustine Campus
|